5 dB to 2 or 1 dB steps Greyed out
1. Problem Description
The dropdown menu used to change step sizes from 5 dB to 2 dB or 1 dB is greyed out and cannot be adjusted. This typically occurs when a test type is configured with the wrong setting in Measure. the Interrupter on test definition is set to Auto Present.
2. Steps to Resolve
-
Open Measure.
-
Go to Pure Tone Measurement.
-
Navigate to Test Type and select the test type where the issue occurs.
-
Right-click and choose Edit Test Definition.
-
In the settings, make sure the test type is set to anything except Auto Present.
-
Click OK to save the changes.
-
Check the dropdown box again and confirm you can now change the dB steps.
3. Expected Outcome
The dropdown menu for adjusting step size (5 dB, 2 dB, or 1 dB) becomes available, and users can freely select their preferred step size.
4. Escalation Criteria
Escalate the issue to support@auditdata.com if:
-
The dropdown remains greyed out after following the steps.
-
The option is unavailable across multiple test types.
-
The software shows errors when editing the Test Definition.
For escalation, gather and provide:
-
Measure version in use.
-
Test type where the issue occurs.
-
Screenshots of the greyed-out dropdown and settings.
-
Steps already attempted.