Fitting Unit Board Not Detected


Quick summary

The Measure device is not recognized by the PC or software. This guide helps you resolve connection issues, muted sound settings, and - if needed - perform a hardware reset to restore board detection.



⚙️ Before you begin

  • Have a spare USB cable available for testing.
  • Make sure you can access the back of the Measure device (you'll need a straightened paperclip or SIM ejector pin for the reset step).
  • Close the Measure software before starting.

Step 1: Disconnect additional devices

Are multiple devices connected to the Primus board?

  • Yes → Disconnect everything except the power cable and USB cable. Retry detection.
  • No → Proceed to Step 2.

Check if the issue is resolved. If yes - one of the disconnected devices was causing a conflict; reconnect them one at a time to identify it. If no - continue to Step 2.


Step 2: Verify cable connections

Are all cables securely and correctly seated?

  • Yes → Proceed to Step 3.
  • No → Reconnect all cables - USB, power, and accessories - making sure each connector is fully inserted.

Check if the issue is resolved. If yes - you're done ✓. If no - continue to Step 3.


Step 3: Try a different USB cable and port

Have you tested with a different USB cable and a different USB port on the PC?

  • Yes → Proceed to Step 4.
  • No → Replace the USB cable with a known working one and connect to a different USB port directly on the PC.

Check if the issue is resolved. If yes - you're done ✓. If no - continue to Step 4.


Step 4: Connect directly to the PC

Are you using a USB extension cable or USB hub?

  • Yes → Disconnect all USB extenders or hubs. Connect the Primus board directly to a USB port on the PC.
  • No → Proceed to Step 5.

Check if the issue is resolved. If yes - you're done ✓. If no - continue to Step 5.


Step 5: Check Windows sound settings

Is Primus muted or disabled in Windows Sound settings?

  • Yes → Take these steps:
  1. Go to Control Panel → Sound.
  2. Right-click PrimusEnable.
  3. Right-click again → Set as Default Device.
  4. Make sure the volume slider is not at zero.

Check if the issue is resolved. If yes - you're done ✓. If no - continue to Step 6. * ❌ No → Proceed to Step 6.


Step 6: Perform a hardware reset

If the board still isn't detected, a hardware reset may restore it to a detectable state.

  1. Disconnect all cables from the board: USB, power supply, and transducers.
  2. Locate the reset pinhole on the back of the board.
  3. Insert a straightened paperclip or SIM ejector pin into the pinhole and hold the reset button down.
  4. While holding the button, connect only the USB cable.
  5. Continue holding for 3 seconds until the LED blinks once per second.
  6. Release the button. The LED should stay on briefly, then turn off.
  7. Open Device Manager (right-click Start → Device Manager) and confirm the board appears.
  8. Launch Primus - it should detect the default firmware and prompt you to update.

Check if the issue is resolved. If yes - you're done ✓. If no - proceed to the escalation path below.


🔺 Escalation path

If none of the above steps resolved the issue, contact Auditdata Support.

Please have ready:

  • Product name and version number
  • Device model and serial number
  • Firmware version
  • Your account name and contact details
  • Which steps above you already tried
  • Exact error messages or codes shown
  • Screenshots if available

✅ Expected outcome

After completing these steps, the board should appear in Device Manager and be recognised by Primus, Measure, or Unity. If a firmware update prompt appears after the reset, complete the update before using the device.


📝 Additional notes

  • The hardware reset (Step 6) clears the board's connection state - it does not erase patient data or calibration files stored on the PC.
  • If Primus prompts for a firmware update after the reset, do not skip it; the update restores the board to its correct operating state.
  • Related articles: No Power or Board Not Detected · Device Unresponsive – Power LED Always On · PRIMUS - Failed Hardware Diagnostics