Quick summary

Calibration readings are unexpectedly low when performing free-field calibration with a REM (Real Ear Measurement) headset. This guide walks through probe validation, calibration sequence, systematic verification, environment setup, hardware checks, and loudspeaker configuration to restore accurate measurements.



⚙️ Before you begin

  • Make sure the testing room is quiet - background noise directly affects calibration accuracy.
  • Have a spare probe set available for hardware testing if needed.
  • Do not move the speaker or patient position during calibration.

Step 1: Verify probe calibration

Is the initial probe calibration valid and are the probes functioning correctly?

  • Yes → Proceed to Step 2.
  • No → Confirm the initial probe calibration is valid and ensure both probes are functioning correctly before continuing.

Check if the issue is resolved. If yes - you're done ✓. If no - continue to Step 2.


Step 2: Follow the correct calibration sequence

Have you followed the correct free-field calibration sequence at the correct distance?

  • Yes → Proceed to Step 3.
  • No → Take these steps in order:
  1. Conduct free-field calibration at a 50 cm distance from the speaker.
  2. Immediately follow with Probe Tube Calibration at the same 50 cm distance.

Check if the issue is resolved. If yes - you're done ✓. If no - continue to Step 3.


Step 3: Calibrate and compare both probes

Have you calibrated and compared results for both left and right probes?

  • Yes → Proceed to Step 4.
  • No → Take these steps:
  1. Perform free-field calibration on the left probe.
  2. Repeat free-field calibration on the right probe.
  3. Compare results between left and right - note any significant differences, as this may indicate a faulty probe.

Check if the issue is resolved. If yes - you're done ✓. If no - continue to Step 4.


Step 4: Optimise the testing environment

Is your testing environment quiet and free from vibrations or external noise?

  • Yes → Proceed to Step 5.
  • No → Take these steps:
  1. Ensure the testing room is completely quiet - turn off ventilation or other noise-generating equipment.
  2. Stabilise probe positioning so it does not shift during calibration.
  3. Minimise external noise interference and use sound-dampening techniques if necessary.

Check if the issue is resolved. If yes - you're done ✓. If no - continue to Step 5.


Step 5: Test with an alternative probe set

Have you tested with a different set of probes to rule out hardware damage?

  • Yes → Proceed to Step 6.
  • No → Try an alternative probe set and repeat the calibration. Check the original probes for physical damage or wear - replace them if worn.

Check if the issue is resolved. If yes - you're done ✓. If no - continue to Step 6.


Step 6: Verify loudspeaker selection and alignment

Is the correct loudspeaker selected and properly aligned in the calibration settings?

  • Yes → Proceed to the escalation path below.
  • No → Take these steps:
  1. Double-check the loudspeaker selection in Tools → Sound Field Calibration settings - it must match the physically connected speaker.
  2. Verify the loudspeaker is correctly aligned and positioned at the right distance and angle.

Check if the issue is resolved. If yes - you're done ✓. If no - proceed to the escalation path below.


🔺 Escalation path

If none of the above steps resolved the issue, contact Auditdata Support.

Please have ready:

  • Specific calibration values observed (e.g. expected vs. actual dB readings)
  • REM headset model and serial number
  • Probe details (model and condition)
  • Description of the testing environment
  • Your account name and contact details
  • Which steps above you already tried
  • Any error messages or unusual behaviour observed

✅ Expected outcome

After completing these steps, free-field calibration should return accurate readings. Measurements should be consistent between left and right probes, and REM headset performance should be reliable.


📝 Additional notes

  • Precise calibration is critical for accurate audiometric testing - always follow the correct sequence (free-field first, then probe tube at the same distance).
  • Environmental factors - including ventilation noise and room reflections - significantly impact calibration accuracy.
  • Regular probe maintenance prevents calibration issues. Replace probe tubes at recommended intervals.
  • Related articles: REM or Free-Field Speaker Calibration Issues or Inaccuracies · Measure - High-Frequency Transducer Not Responding · Measure – Transducer Connection Issues and Incorrect Free Field Calibration