Measure – Transducer Connection Issues and Incorrect Free Field Calibration
- Product(s): Measure
- Applies to version: All current versions
- Audience: External – Customers
- Difficulty: Advanced
- Estimated resolution time: 20–30 minutes
- Last updated: 2026-06-27
- Article type: Troubleshooting guide
Quick summary
Measure is intermittently not detecting transducers, or free-field calibration levels are incorrect. This guide walks you through transducer reconnection, free-field recalibration, and a manual hardware reset if needed.
⚙️ Before you begin
- Make sure Measure is open and running.
- Have a thin pin (e.g. a straightened paperclip) available in case a manual hardware reset is needed.
Step 1: Reconnect transducers and run hardware diagnostics
Are all transducers shown as connected in Measure?
- ✅ Yes → Proceed to Step 2.
- ❌ No → Take these steps:
- Disconnect each transducer and reconnect it firmly.
- If available, try alternative cables.
- Run the Hardware Diagnostic Test from within Measure.
→ Check if the issue is resolved. If yes - you're done ✓. If no - continue to Step 2.
Step 2: Recalibrate free-field speakers
Is the free-field speaker output calibrated correctly at 80 dB for pure tone audiometry?
- ✅ Yes → Proceed to Step 3.
- ❌ No → Take these steps:
- Recalibrate the free-field speakers to 80 dB for pure tone audiometry.
-
Retest the output:
- If soft or no noise is present, recalibrate again.
- If there is no sound at all, test on a different PC and verify the REM speaker settings.
→ Check if the issue is resolved. If yes - you're done ✓. If no - continue to Step 3.
Step 3: Perform a manual hardware reset
Have all previous steps failed to resolve the issue?
- ✅ Yes → Perform a manual hardware reset of the audiometer board:
- Disconnect all cords, transducers, and the power supply from the board.
- Locate the small reset hole on the back of the board.
- Insert a thin pin into the reset hole until you feel a light click.
- Hold the pin in place and reconnect the power supply while still holding.
- Wait for the device to blink, then release the pin - do not disconnect the power.
- Once the board has reset, recalibrate the free-field speakers to 80 dB for pure tone audiometry.
→ Check if the issue is resolved. If yes - you're done ✓. If no - proceed to the escalation path below. * ❌ No → The issue is resolved - no further action needed.
🔺 Escalation path
If none of the above steps resolved the issue, contact Auditdata Support.
Please have ready:
- Device model and serial number
- Device firmware version
- Software version
- Your account name and contact details
- Which steps above you already tried
- Any error messages or unusual behaviour observed
✅ Expected outcome
After completing these steps, transducers should be consistently detected and shown as connected in Measure. Free-field speakers should emit correct sound levels at 80 dB, detectable via REM probe.
📝 Additional notes
- Any Measure device returned from repair must be recalibrated for free-field speakers to 80 dB for pure tone audiometry before use.
- Any board reset - manual or software - also requires free-field speaker recalibration.
- Related articles: REM or Free-Field Speaker Calibration Issues or Inaccuracies · PRIMUS - Failed Hardware Diagnostics · Fitting/HIT Unit Not Connecting or Functioning Correctly
Need more help? Visit the Auditdata Help Centre or contact our support team.