Quick summary
Transducers or loudspeakers in Measure cannot reach their maximum output level. This is typically caused by the power supply not being connected, transducer or loudspeaker hardware limitations, or a calibration issue. This guide walks you through the most common checks and fixes.
⚙️ Before you begin
- Check that all cables are firmly connected at both ends.
- Confirm which transducer or loudspeaker is affected.
Step 1: Check the power supply
Is the Measure power supply plugged in and functioning correctly?
- ✅ Yes → Proceed to Step 2.
- ❌ No → Take these steps:
- Plug in the Measure power supply securely.
- Check that the power LED turns on and Measure is receiving power.
- Try a different power outlet if the LED does not turn on.
→ Check if the issue is resolved. If yes - you're done ✓. If no - continue to Step 2.
Step 2: Check high-frequency playback and calibration
Can the transducer or loudspeaker play high-frequency sounds at all?
- ✅ Yes → The hardware can reach high output at some frequencies - the limitation may be frequency-specific. Proceed to the escalation path below to investigate further.
- ❌ No → This may be a hardware limitation or a calibration issue:
- Check whether the transducer or loudspeaker is specified to support the frequency range you are testing. Refer to the transducer's datasheet or the Measure compatibility list.
- If the hardware should support the range, contact your calibration provider - the transducer or loudspeaker may need recalibration.
→ Check if the issue is resolved. If yes - you're done ✓. If no - proceed to the escalation path below.
🔺 Escalation path
If the transducer or loudspeaker still cannot reach full output after completing the steps above, contact Auditdata Support.
Please have ready:
- Device model and serial number
- Transducer or loudspeaker model
- Measure software version
- Device firmware version
- Your account name and contact details
- Which steps above you already tried
- Any error messages or unusual behaviour observed
✅ Expected outcome
After completing these steps, the transducer or loudspeaker should play at its full rated output level, including high frequencies where supported by the hardware.
📝 Additional notes
- Transducer output limits are defined by the hardware specifications - Measure cannot exceed the physical limitations of the connected device.
- Keep Measure updated to the latest version to ensure accurate output management.